Actual | YOCON automatic nucleic acid extraction workstation can effectively prevent cross-contamination!

Yocon Market

2021-10-20

Company Events


“Accurate, fast, convenient and secure”

This is a prioritization of our requirements for the workstation.

The workstation has to be put into operation.

One of the biggest concerns that customers often have about such product is that -

Whether the equipment will produce a false positive result (cross-contamination of different samples)?

To make customers more assured, we have tested the FE4800 automatic nucleic acid extraction workstation, using real virus samples (purchased from national institutions through regular channels, without infectivity), and have made the first success!


I. Background

What is the possible cause for cross-contamination of samples?

--Too short distance between sampling tubes and between different deep-well plates, and inevitable aerosol produced during liquid transfer by equipment. Cross-contamination of samples may occur if aerosols from one sample fall into another. Our equipment processes 8 samples simultaneously, and it is normal for customers to have doubts.


Why should we prevent such cross-contamination?

--Not all cross-contaminations have serious consequences, but cross-contamination in the molecular biology does produce extremely serious consequences because the back-end testing principle is multiple replication of the initial sample. Based on the current CT value of 35 as the cut-off value for a positive or negative diagnosis during the clinical test, the initial sample should be replicated 235 times, that is, 64.0 billion times. It is amazing that one nucleic acid can be amplified to 64.0 billion nucleic acids. Therefore, whenever a sample is contaminated by another one, result errors are inevitable.


2.Objective

To systematically evaluate the performance of YOCON automatic sample processing and nucleic acid extraction workstation and to verify that there is no aerosol contamination during normal extraction by the equipment.


3.Design

1、 A large number of strongly positive samples and negative samples were arranged in a staggered manner for replicate tests of 3 batches to test whether the extraction equipment would lead to cross-contamination under extreme conditions.

2、 A total of 48 samples were extracted for one batch, including 27 virus-containing samples and 21 virus-free samples.

3、 The virus concentration of the positive samples was 109 copies/mL (refer to the upper limit of actual concentration of positive clinical samples).

4、 Virus-containing samples and blank samples were arranged in a staggered manner, with positive samples around negative sample (negative-6 sample as shown below) and negative samples around positive sample (2-1 sample as shown below), as detailed below:


Arrangement of negative and positive samples.

Negative

Negative

Negative

Positive

Negative

Positive

Negative

Positive

Negative

Positive

Negative

Positive

Negative

Positive

Negative

Positive

Negative

Negative

Negative

Positive

Positive

Negative

Positive

Negative

Positive

Positive

Positive

Positive

Positive

Positive

Positive

Negative

Positive

Negative

Positive

Positive

Positive

Negative

Positive

Positive

Positive

Positive

Positive

Positive

Negative

Negative

Negative

Negative


4.Procedures

1、 The virus mixture was prepared.

2、 The virus samples were added. The virus mixture was added to the sampling tube in a 1:9 volume ratio. A total of 27 virus-containing sampling tubes were set, marked with serial numbers, and recorded. One swab was broken and put into each sampling tube. The tubes were shaken well and mixed well.

3、 Preparation of the reagents required for the instrument. A total of three reagents were required, that is, 15.36 mL of a mixture of isopropanol and magnetic beads (containing 14.4 mL of isopropanol and 960 μL of magnetic beads), 38.4 mL of purification solution and 4.8 mL of cleaning solution.

4、 27 tubes were inserted into the pre-set wells from left to right and top to bottom, and the negative control tube without virus sample were inserted into the remaining 21 wells to start the program for extraction.

5、 The reverse transcription fluorescence quantitative PCR assay was carried out for 21 negative samples, and one well containing viral nucleic acid was set as positive control.

6、 The replicate tests were carried out for 3 batches.


5.Results

1. The expansion curves of each group for batch 1 are as follows:

擴(kuò)增曲線1.jpg

Note: Only the positive control group had an expansion curve, while the negative control and others showed no expansion.


Statistics:

Group

Ct value

Group

Ct value

Group

Ct value

Negative-1

NoCt

Negative-8

NoCt

Negative-15

NoCt

Negative-2

NoCt

Negative-9

NoCt

Negative-16

NoCt

Negative-3

NoCt

Negative-10

NoCt

Negative-17

NoCt

Negative-4

NoCt

Negative-11

NoCt

Negative-18

NoCt

Negative-5

NoCt

Negative-12

NoCt

Negative-19

NoCt

Negative-6

NoCt

Negative-13

NoCt

Negative-20

NoCt

Negative-7

NoCt

Negative-14

NoCt

Negative-21

NoCt

Positive control

27.95

Negative control

NoCt




2. The expansion curves of each group for batch 2 are as follows:

擴(kuò)增曲線2.jpg

Note: Only the positive control group had an expansion curve, while the negative control and others showed no expansion.


Statistics:

Group

Ct value

Group

Ct value

Group

Ct value

Negative-1

NoCt

Negative-8

NoCt

Negative-15

NoCt

Negative-2

NoCt

Negative-9

NoCt

Negative-16

NoCt

Negative-3

NoCt

Negative-10

NoCt

Negative-17

NoCt

Negative-4

NoCt

Negative-11

NoCt

Negative-18

NoCt

Negative-5

NoCt

Negative-12

NoCt

Negative-19

NoCt

Negative-6

NoCt

Negative-13

NoCt

Negative-20

NoCt

Negative-7

NoCt

Negative-14

NoCt

Negative-21

NoCt

Positive control

28.55

Negative control

NoCt




3. The expansion curves of each group for batch 3 are as follows:

擴(kuò)增曲線3‘’.jpg

Note: Only the positive control group had an expansion curve, while the negative control and others showed no expansion.


Statistics:

Group

Ct value

Group

Ct value

Group

Ct value

Negative-1

NoCt

Negative-8

NoCt

Negative-15

NoCt

Negative-2

NoCt

Negative-9

NoCt

Negative-16

NoCt

Negative-3

NoCt

Negative-10

NoCt

Negative-17

NoCt

Negative-4

NoCt

Negative-11

NoCt

Negative-18

NoCt

Negative-5

NoCt

Negative-12

NoCt

Negative-19

NoCt

Negative-6

NoCt

Negative-13

NoCt

Negative-20

NoCt

Negative-7

NoCt

Negative-14

NoCt

Negative-21

NoCt

Positive control

29.72

Negative control

NoCt



6.Conclusions

During the extraction tests for the same batch, negative samples and positive samples were arranged in a staggered manner, and contamination was not found in negative sample wells, but found in positive control well, indicating that this equipment can effectively prevent inter-sample cross-contamination.


Words in the end:

We choose to be one of the few automatic nucleic acid workstation suppliers to offer measured data of samples using our automatic nucleic acid workstation. We will continue to test multiple batches of samples in the future to ensure that the equipment meets the customer needs. Let’s look forward to it!


熟女高潮一区二区 | 潮吹 的搜索结果 - 91n | 亚洲乱熟乱熟女一区二区 | 久久无码人妻一区二区三区午夜免费 | 少妇搡BBBB搡BBB搡野外 | 一本大道HEYZO无码中文字幕 | 啊啊亚洲无吗视频免费 | 天美精品一区二区三区, | 14女人毛片A片免费 91高清无码无套内射 | 亚日韩在线观看了 | 97人人添人人澡人人爽亚洲正品 | 国产乱子伦精品无码码专区 | 少妇搡BBBB搡BBB搡打电话 | 午夜福利1000 | 波多野结衣在线播放 | 草莓视频在线52x导航 | 国产女人下身毛毛A片 | 欧美人与性囗牲恔配视频 | 无码高清视频在线观看 | 欧美精品v欧洲精品黑人 | 无码人妻一区二区三区线花季传件 | 国产有大有粗有黄的视频 | 亚洲男人的天堂一区在线免费播放 | ,亚洲人成毛片在线播放 | 欧美老妇 日韩 | 岛国 xxxxxx1 高清在线 | 国产校花久色在线观看 | 一级全黄录像免费观看 | 人人操人人爱人人爽 | 四川BBB又粗又硬又大 | 特级西西444www无码视频免费看 | 在线免费观看黄色视频 | 欧美喷潮喷水失禁合集 | 色先锋影音A∨资源网 | 成人无码精品一区二区 | 国产成人片在线播放 | 可以免费观看的一级毛片 | 亚洲无码中文字幕在线观看 | 日本欧美久久久久免费播放网 | 热久久一二三四五六馆an | 免费黄污高清无码网站 |